R&S ESU
Measuring Harmonics with Frequency Sweeps
5.26
Quick Start Guide 1302.6163.62 - 04
6. Make the step size for the center frequency equal to the signal frequency.
➢
Press the
FREQ
key.
➢
Press the
CF STEPSIZE
softkey and press the
=MARKER
softkey in the submenu.
The step size for the center frequency is now equal to the marker frequency.
7. Set the center frequency to the 2
nd
harmonic of the signal
➢
Press the
FREQ
key.
➢
Press the up arrow key once.
The center frequency is set to the 2
nd
harmonic.
8. Place the delta marker on the 2
nd
harmonic.
➢
Press the
MKR->
key.
➢
Press the
PEAK
softkey.
The delta marker jumps to the maximum of the 2
nd
harmonic. The displayed level result
is relative to the reference point level (= fundamental wave level).
Fig. 5-12 Measuring the level difference between the fundamental wave (= reference
point level) and the 2
nd
harmonic
Содержание esu
Страница 1: ...R S ESU EMI Test Receiver Quick Start Guide Quick Start Guide Test Measurement 1302 6163 62 04 ...
Страница 26: ...R S ESU Contents 6 Quick Start Guide 1302 6163 62 04 ...
Страница 28: ......
Страница 32: ...R S ESU 0 4 Quick Start Guide 1302 6163 62 04 ...
Страница 88: ...R S ESU Preparing for Use Operating System Properties 2 38 Quick Start Guide 1302 6163 62 04 ...
Страница 104: ...R S ESU Basic Operation Setting Parameters 4 12 Quick Start Guide 1302 6163 62 04 ...
Страница 175: ...Quick Start Guide 1302 6163 62 04 6 33 R S ESU LAN Interface Manual Operation of the R S ESU with XP Remote Desktop ...
Страница 178: ...R S ESU LAN Interface Manual Operation of the R S ESU with XP Remote Desktop 6 36 Quick Start Guide 1302 6163 62 04 ...
Страница 201: ...Quick Start Guide 1302 6163 62 04 R S ESU Appendix Appendix ...
Страница 202: ...Quick Start Guide 1302 6163 62 04 R S ESU Appendix ...