R&S ESU
Measuring a Sinusoidal Signal
5.20
Quick Start Guide 1302.6163.62 - 04
Fig. 5-8
Measurement of the frequency with the frequency counter
Obtaining a correct result when measuring the frequency with the
internal frequency counter requires that an RF sinusoidal signal or a
spectral line be present. The marker must be located more than 25 dB
above the noise level to ensure that the specified measurement
accuracy is adhered to.
For bandwidths between 200 kHz and 10 MHz, the time for measuring
the frequency is inversely proportional to the selected resolution, i.e. a
resolution of 1 Hz requires a gate time of 1 second for the counter.
For digital bandwidths of less than 300 kHz, the time for measuring the
frequency is always approx. 30 ms and is not affected by the resolution
that is set.
Thus, when measuring the frequency of a sinusoidal carrier that has
high resolution, set the resolution bandwidth to 100 kHz or less.
Содержание esu
Страница 1: ...R S ESU EMI Test Receiver Quick Start Guide Quick Start Guide Test Measurement 1302 6163 62 04 ...
Страница 26: ...R S ESU Contents 6 Quick Start Guide 1302 6163 62 04 ...
Страница 28: ......
Страница 32: ...R S ESU 0 4 Quick Start Guide 1302 6163 62 04 ...
Страница 88: ...R S ESU Preparing for Use Operating System Properties 2 38 Quick Start Guide 1302 6163 62 04 ...
Страница 104: ...R S ESU Basic Operation Setting Parameters 4 12 Quick Start Guide 1302 6163 62 04 ...
Страница 175: ...Quick Start Guide 1302 6163 62 04 6 33 R S ESU LAN Interface Manual Operation of the R S ESU with XP Remote Desktop ...
Страница 178: ...R S ESU LAN Interface Manual Operation of the R S ESU with XP Remote Desktop 6 36 Quick Start Guide 1302 6163 62 04 ...
Страница 201: ...Quick Start Guide 1302 6163 62 04 R S ESU Appendix Appendix ...
Страница 202: ...Quick Start Guide 1302 6163 62 04 R S ESU Appendix ...