R&S ESU
Instrument Functions
Analyzer Mode
4.122
Operating Manual 1302.6163.12 - 03
Remote command:
SWE:COUN 64
SWEEP
POINTS
The
SWEEP POINTS
softkey selects the number of measurement samples
acquired during a sweep.
The following numbers of points per sweep are available: 155, 201, 301, 313, 401,
501, 601, 625 (default), 701, ..., 30001
An increment of 100 is possible for number of points
≥
201.
Additionally the numbers 1251 and 1999 are available.
Remote command:
SWE:POIN 501
SGL SWEEP
DISP OFF
The
SGL SWEEP DISP OFF
softkey deactivates the display while a single sweep is
being performed. Once the sweep has been completed, the trace is shown.
Remote command:
INIT:DISP OFF
4.8.6
Triggering the Sweep – TRIG Key
The
TRIG
key opens a menu for selection of the various trigger sources, trigger
polarity and external gate function. The active trigger mode is indicated by highlight-
ing the corresponding softkey.
For video trigger, a trigger threshold can be entered, which is represented in the dia-
gram as a horizontal line.
To indicate that a trigger mode other than
FREE RUN
has been set, the enhance-
ment label
TRG
is displayed on the screen. If two windows are displayed, TRG
appears next to the appropriate window.
TRIG
The number of sweeps set in the
TRACE
menu is the same as that in the
SWEEP
menu.
If SINGLE SWEEP is selected, the measurement stops after the selected number
of sweeps has been performed.
The autopeak detector will be disabled while the number of points per sweep is set
to another value than the default.
!
Содержание ESU Series
Страница 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual...
Страница 25: ......
Страница 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03...
Страница 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03...
Страница 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03...
Страница 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03...
Страница 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03...
Страница 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03...
Страница 821: ...R S ESU Remote Control Programming Examples More Complex Programming Examples 7 32 Operating Manual 1302 6163 12 03...
Страница 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03...
Страница 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU...
Страница 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03...