[SENSe:]SWEep:COUNt
<SweepCount>
This command defines the number of scans performed during a single sweep.
Parameters:
<SweepCount>
Range:
0 to 32767
*RST:
0
Example:
SWE:COUN 64
Sets the number of sweeps to 64.
INIT:CONT OFF
Switches to single-sweep mode.
INIT;*WAI
Starts a sweep and waits for its end.
Manual operation:
See
[SENSe:]SWEep:COUNt:CURRent
This command queries the number of scans that have started in single scan mode with
more than one scan (
SENSe:SWEep:COUNt
>
1).
Example:
SWE:COUNt 64
sets scan count to 64
INIT:CONT OFF
switches to single scan mode
INIT
starts a scan (without waiting for the scan end!)
SWE:COUN:CURR?
queries the number of started scans
Manual operation:
See
Remote Commands in Receiver Mode