![R&S ENV216 Скачать руководство пользователя страница 5](http://html1.mh-extra.com/html/rands/env216/env216_user-manual_3137419005.webp)
Safety Instructions for LISNs
R&S
®
ENV216
5
User Manual 5201.6693.14 ─ 04
The ground connection must remain until the LISN has been disconnected from
the power source.
If the LISN has an auxiliary voltage: Secure the test setup with a second protec-
tive ground. The second ground connection protects you from live parts if the
ground connection of the auxiliary cable fails.
Connecting to power
The product is an overvoltage category II product and must be connected to a
fixed installation used to supply energy-consuming equipment such as household
appliances and similar loads. Be aware that electrically powered products have
risks, such as electric shock, fire, personal injury or even death.
Keep to the protective measures given by VDE
0100-410 (IEC60634-4-41) while
working with the LISN.
Take the following measures for your safety:
●
Use double-insulated connecting cables with characteristics suitable for the
application:
– Appropriate minimum cross-section.
– Appropriate minimum voltage rating.
Select the cable based on the maximum power consumption of the device
under test, the fuse rating of the building installation and the cable length.
When installing the supplied cable sockets, observe the assembly instructions
and the requirements of the manufacturer.
●
By design, LISNs do not have a fuse in the measuring circuit. Therefore, you
have to make sure that the operating circuit between the power supply and
the LISN's power connector is fused correctly.
●
Only use the power cable delivered with the product. It complies with country-
specific safety requirements.
●
Only use intact cables and route them carefully so that they cannot be dam-
aged. Check the power cables regularly to ensure that they are undamaged.
Also ensure that nobody can trip over loose cables.
●
Only connect the product to a power source with a maximum fuse protection
according to the datasheet.
●
Ensure that you can disconnect the product from the power source at any
time. Pull the power plug to disconnect the product. The power plug must be
easily accessible. If the product is integrated into a system that does not meet
these requirements, provide an easily accessible circuit breaker at the system
level.