Getting Started 4072.8432.02 ─ 04
a) "MINIMUM": the minimum of the FFT trace over time is displayed.
b) "MAXIMUM": the maximum of the FFT trace over time is displayed.
c) "AVERAGE": the average of the FFT trace over time is displayed.
d) "CLRWRITE": the instantaneous FFT is displayed without averaging.
3. The Demodulation Frequency softkey allows a small portion of the displayed IF
Span to be selected for demodulation.
a) Select the "DEMOD FREQ" softkey to highlight it.
b) With the softkey in (3a) highlighted, follow procedure (1) to change the
demodulation frequency within the IF span.
c) The Mid Frequency will not change.
4. The Receiver Control Keys (
chapter 2.1, "Front-Panel Tour"
Receiver Control Keys) allow changes to various receiver parameters.