R&S AFQ100A
Display
1401.3078.62 3.9
E-3
Info Window - Display
A few operating states and the current message are displayed in the info line (see also chapter 9 "
").
The info window with a list of current permanent messages and a detailed description of each message
can be opened by clicking on the
Info
softkey.
The upper section of the info window contains a list of all current permanent messages in the order of
their occurrence, i.e. the most recent message is displayed first. In the lower section of the window,
additional information on the highlighted message is displayed.
A history of all messages that have occurred since instrument switch-on can be called with the
History
softkey. The most recent message is displayed first.
Info line
List of current
messages with short
message text.
Detailed description
for highlighted
message
Softkeys
The messages are color-coded according to their level. Device-specific messages are red, information
and remote control errors are black. The level is also indicated in the
Lev
column (Err, Sys or Info).
Column
SCPI
indicates the SCPI error code.