Configuration for All Measurement Types
R&S
®
NRX
64
User Manual 1178.5566.02 ─ 07
Remote command:
TRIGger<undef>:ALL:DELay[:VALue]
TRIGger<Measurement>[:CHANnel<Channel>]:DELay[:VALue]
Dropout ← Trigger Advanced
If the trigger level has been underrun with internal triggering, it is sometimes useful to
wait some time (drop-out time) before retriggering. The measurement of unwanted sig-
nal components can thus be suppressed. See
Remote command:
TRIGger<Measurement>[:CHANnel<Channel>]:DTIMe
Holdoff ← Trigger Advanced
Sets the hold-off time, see
Chapter 6.3.4, "Hold-Off Time"
Remote command:
TRIGger<Measurement>[:CHANnel<Channel>]:HOLDoff
Hysteresis ← Trigger Advanced
Sets the hysteresis. A trigger event occurs, if the trigger level:
●
Falls below the set value on a rising slope.
●
Rises above the set value on a falling slope.
Thus, you can use this setting to eliminate the effects of noise in the signal for the edge
detector of the trigger system.
Remote command:
TRIGger<Measurement>[:CHANnel<Channel>]:HYSTeresis
Specific Trigger
Not available for each sensor type. Groups the specific trigger settings.
Jitter Suppression ← Specific Trigger
Defines the method how to cope with the misalignment between the trigger event and
the sample point.
"Compensate"
Compensation means resampling of trace result.
"Measure"
Does not perform resampling, but stores the measured trigger jitter.
Remote command:
TRIGger<Measurement>[:CHANnel<Channel>]:JITTer:METHod
Trigger Master State ← Specific Trigger
Enables or disables the power sensor as trigger master. If enabled, the power sensor
outputs a digital trigger signal in sync with its own trigger event. The trigger signal is
output at the port selected under
Triggering