Remote Control Commands
R&S
®
NRX
223
User Manual 1178.5566.02 ─ 07
Sets the termination resistance of the external trigger signal that is supplied at the trig-
ger input/output of the power sensor. Choose a setting that fits the impedance of the
trigger source to minimize reflections on the trigger signals.
Suffix:
<Measurement>
.
1 to 8
Measurement
<Channel>
1 to 2
1 = primary sensor, 2 = secondary sensor
<Port>
1 to 2
Power sensor ports; 1 = USB port, 2 = trigger I/O connector
Parameters:
<impedance>
HIGH | LOW
HIGH
~
10
k
Ω
LOW
50
k
Ω
*RST:
HIGH
Manual operation:
See
TRIGger<undef>:ALL:HOLDoff
<holdoff>
TRIGger<Measurement>[:CHANnel<Channel>]:HOLDoff
<holdoff>
Sets the hold-off time, see
Chapter 6.3.4, "Hold-Off Time"
The command is not fully R&S NRP2 compatible. If port
1 is assigned to measure-
ment
1 and so on, the command works. If port
1 is assigned to measurement
2, for
example, enable the R&S NRP2 emulation using
"NRP2"
.
Suffix:
<Measurement>
.
1 to 8
Measurement
<Channel>
1 to 2
1 = primary sensor, 2 = secondary sensor
Parameters:
<holdoff>
Range:
0.0 to 10.0
*RST:
0.0
Default unit: s
Manual operation:
See
TRIGger<undef>:ALL:HYSTeresis
<hysteresis>
TRIGger<Measurement>[:CHANnel<Channel>]:HYSTeresis
<hysteresis>
Sets the hysteresis. A trigger event occurs, if the trigger level:
●
Falls below the set value on a rising slope.
●
Rises above the set value on a falling slope.
Measurement Settings and Results