Configuration for All Measurement Types
R&S
®
NRX
60
User Manual 1178.5566.02 ─ 07
●
Measuring
The power sensor is measuring data. It remains in this state during the measure-
ment. When the measurement is completed, it exits this state immediately.
6.3.2
Trigger Sources
The possible trigger conditions and the execution of a trigger depend on the selected
trigger mode and trigger source.
If the signal power exceeds or falls below a reference level set by the trigger level, the
measurement is started after the defined delay time. Waiting for a trigger event can be
skipped.
Trigger source
Description
Remote commands to initiate the measurement
"Hold"
Waits for a trigger event. Press [1Trig] to trigger
the measurement.
Depending on the sensor type, the trigger is exe-
cuted by the trigger bus or by remote command.
See the user manual of the power sensor for
details.
TRIGger<Measurement>[:IMMediate]
"Immediate"
Measures immediately, does not wait for trigger
condition.
-
"Internal"
Uses the input signal as trigger signal.
TRIGger<Measurement>[:IMMediate]
"Internal
A" , "Inter-
nal
B" , "Internal
C" ,
"Internal
D"
Receives the trigger signal from the trigger mas-
ter. The trigger master is the sensor connected to
port
A, B, C, or D. See
TRIGger<Measurement>[:IMMediate]
"External"
Uses the external trigger signal that is supplied at
the Trig In / Out
2 connector. See
TRIGger<Measurement>[:IMMediate]
"External
2"
Requires a power sensor with a trigger input/
output.
Uses the external trigger signal that is supplied at
the trigger input/output of the power sensor.
TRIGger<Measurement>[:IMMediate]
"Sensor Check
Source"
Requires the sensor check source (R&S NRX-B1)
option.
If enabled, the sensor check source (R&S NRX-
B1) sends trigger signals using the internal trigger
bus. See
TRIGger<Measurement>[:IMMediate]
"Bus (*TRG)"
Waits for a trigger event. Press [1Trig] to trigger
the measurement.
Depending on the sensor type, the trigger is exe-
cuted by the trigger bus or by remote command.
See the user manual of the power sensor for
details.
TRIGger<Measurement>[:IMMediate]
Triggering
www.allice.de
Allice Messtechnik GmbH