Remote Control Commands
R&S
®
NRX
228
User Manual 1178.5566.02 ─ 07
Parameters:
<port>
INTernal | EXT2 | EXTernal2
*RST:
INTernal
Manual operation:
See
TRIGger<undef>:ALL:SYNChronize[:STATe]
<state>
TRIGger<Measurement>[:CHANnel<Channel>]:SYNChronize[:STATe]
<state>
Usually used if
TRIGger<Measurement>[:CHANnel<Channel>]:MASTer[:
ON
is set.
If enabled, blocks the external trigger bus as long as the power sensor remains in the
measurement state. Thus, ensures that a new measurement is only started after all
power sensors have completed their measurements.
Make sure that the number of repetitions is the same for all power sensors involved in
the measurement. Otherwise, the trigger bus is blocked by any power sensor that has
completed its measurements before the others and has returned to the idle state.
The command is not fully R&S NRP2 compatible. If port
1 is assigned to measure-
ment
1 and so on, the command works. If port
1 is assigned to measurement
2, for
example, enable the R&S NRP2 emulation using
"NRP2"
.
Suffix:
<Measurement>
.
1 to 8
Measurement
<Channel>
1 to 2
1 = primary sensor, 2 = secondary sensor
Parameters:
<state>
ON | OFF
*RST:
OFF
Manual operation:
See
TRIGger<undef>:ALL[:IMMediate]
TRIGger<Measurement>[:IMMediate]
Starts a measurement.
Suffix:
<Measurement>
.
1 to 8
Measurement
Usage:
Event
13.5.3
Selecting the Measurement
Before starting a measurement, select the measurement type.
Measurement Settings and Results