Remote Control Commands
R&S
®
NRP18S-xx
81
User Manual 1178.3686.02 ─ 05
[SENSe<Sensor>:]TRACe:TIME
<time>
Sets the trace length, time to be covered by the trace sequence. This time period is
divided into a number of equal intervals, in which the average power is determined.
The number of intervals equals the number of trace points, which is set with the com-
mand
Parameters:
<time>
Range:
10.0e-6 to 3.0
*RST:
0.01
Default unit: Seconds
8.8
Configuring Basic Measurement Parameters
The following section describes the settings common for several measurement modes.
8.8.1
Configuring Auto Averaging
Describes the commands for automatic averaging in the continuous average and burst
measurements.
Remote commands:
..................................................................................81
[SENSe<Sensor>:]AVERage:COUNt:AUTO
........................................................................82
[SENSe<Sensor>:]AVERage:COUNt:AUTO:MTIMe
.............................................................82
[SENSe<Sensor>:]AVERage:COUNt:AUTO:NSRatio
...........................................................82
[SENSe<Sensor>:]AVERage:COUNt:AUTO:RESolution
.......................................................83
[SENSe<Sensor>:]AVERage:COUNt:AUTO:SLOT
.............................................................. 83
...................................................................................83
[SENSe<Sensor>:]AVERage:COUNt:AUTO:TYPE
.............................................................. 84
[SENSe<Sensor>:]AVERage:TCONtrol
.............................................................................. 84
[SENSe<Sensor>:]AVERage[:STATe]
................................................................................. 84
[SENSe<Sensor>:]AVERage:COUNt
<count>
Sets the number of readings that are averaged for one measured value. The higher the
count, the lower the noise, and the longer it takes to obtain a measured value.
Average count is often also called averaging factor, but it designates the same thing,
i.e the number of measured values that have to be averaged for forming the measure-
ment result.
Averaging is only effective, if
[SENSe<Sensor>:]AVERage[:STATe]
ON
is set.
Parameters:
<count>
Range:
1 to 65536
*RST:
4
Example:
AVER:COUN 1
Configuring Basic Measurement Parameters