![R&S 1424.6721.02 Скачать руководство пользователя страница 140](http://html1.mh-extra.com/html/rands/1424-6721-02/1424-6721-02_user-manual_774880140.webp)
Remote Control Basics
R&S
®
NRP18S-xx
140
User Manual 1178.3686.02 ─ 05
10.3.5
Device Status Register
0
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
Sensor error summary
Sensor error
Sensor error
Sensor error
Sensor error
0
0
Legacy locked
Reference PLL locked
0
0
0
0
0
0
0
+
Figure 10-3: Device status register
Querying the register:
●
●
Querying the static errors:
●
Table 10-5: Used device status bits and their meaning
Bit no.
Short description
Bit is set if
0
Sum of SERR bits
Sum/combination of SERR bits 1 to 4.
1
SERR measurement not possi-
ble
2
SERR erroneous results
Static error exists. The measurement result is possibly incor-
rect.
3
SERR warning
Static error exists. Status LED of the power sensor is slowly
flashing red.
4
SERR critical
Critical static error exists. Status LED of the power sensor is
fast flashing red.
Status Reporting System