![R&S 1424.6721.02 Скачать руководство пользователя страница 65](http://html1.mh-extra.com/html/rands/1424-6721-02/1424-6721-02_user-manual_774880065.webp)
Remote Control Commands
R&S
®
NRP18S-xx
65
User Manual 1178.3686.02 ─ 05
FORMat:SREGister
<sregister>
Specifies which format is used for the return value of
Parameters:
<sregister>
ASCii | HEXadecimal | OCTal | BINary
*RST:
ASCii
Example:
FORM:SREG ASC
FORMat[:DATA]
[<data,length>, <length>]
Specifies how the R&S NRP18S-xx sends the numeric data to the controlling host/
computer.
Parameters:
<data,length>
<REAL,32 | 64>
REAL
Block of binary values, 32-bit or 64-bit each; so-called "SCPI
definite length block"
32 | 64
32-bit or 64-bit
If you omit the length, the R&S NRP18S-xx sets the last used
length.
Example for
REAL,32
format:
#6768000....<binary float values>....
Example for
REAL,64
format:
#71536000....<binary float values>....
<data[,length]>
<ASCii[,0 to 12]>
ASCii
List of comma separated, readable values.
[,0 to 12]
Defines the number of decimal places.
The reset value 0 does not restrict the number of decimal pla-
ces.
Example for
ASCii,4
format:
1.2938e-06, -4.7269e-11, ...
*RST:
ASCii,0
8.6.3
Retrieving Results
After performing the measurement, you can query the measurement results with a
command from the
FETCh
subsystem.
Remote commands:
.............................................................................................66
FETCh<Sensor>:ARRay[:POWer][:AVG]?
.......................................................................... 67
FETCh<Sensor>[:SCALar][:POWer][:AVG]?
........................................................................67
Configuring and Retrieving Results