Command reference
R&S
®
ZNA
1350
User Manual 1178.6462.02 ─ 20
<ImaginaryPart>
Imaginary part of the port impedance; may be omitted to define
a real impedance.
Range:
- 10 MΩ to 10 MΩ
*RST:
0 Ω (real impedances)
Default unit: Ohm
Example:
*RST; :CALC:PAR:DEL 'TRC1'
Reset the analyzer and delete the (default) trace for channel no.
1.
SOUR:LPOR1 1,2
Combine the physical ports no. 1 and 2 to define the balanced
(logical) port no. 1.
LPOR1:ZDIF 27, 2
Specify a complex differential mode reference impedance of
27Ω + j * 2Ω for the defined port.
Manual operation:
See
"Single Ended Mode / Common Mode / Differential Mode"
[SENSe<Ch>:]LPORt<LogPt>:ZDEFault[:STATe]
<arg0>
Allows to toggle between default and renormalized reference impedance(s) for logical
port <LogPt>.
Suffix:
<Ch>
.
Channel number
<LogPt>
Logical port number
Parameters:
<arg0>
ON (1): Use default impedance(s)
OFF (0): Use the redefined impedances defined via
[SENSe<Ch>:]PORT<PhyPt>:ZREFerence
or via
[SENSe<Ch>:]LPORt<LogPt>:ZCOMmon
[SENSe<Ch>:]LPORt<LogPt>:ZDIFferent
logical ports
Manual operation:
See
7.3.14.17
[SENSE:]NFIGure...
The
[SENSE:]NFIGure...
commands define properties of a
.
[SENSe<Ch>:]NFIGure:CALibration:NTIMe
.................................................................... 1351
[SENSe<Ch>:]NFIGure:CALibration:GTIMe
.................................................................... 1351
.....................................................................................1351
......................................................................................1352
[SENSe<Ch>:]NFIGure:HARMonic[:STATe]
.................................................................... 1352
[SENSe<Ch>:]NFIGure:HARMonic[:MINFrequency]
........................................................ 1352
[SENSe<Ch>:]NFIGure:HARMonic[:MAXFrequency]
....................................................... 1352
[SENSe<Ch>:]NFIGure:HARMonic[:MAXimumorder]
....................................................... 1353
SCPI command reference