Command reference
R&S
®
ZNA
1353
User Manual 1178.6462.02 ─ 20
Parameters:
<FrequencyLimit>
Frequency limit at which the noise figure sideband correction is
used.
Range:
10 MHz | 9 GHz
Default unit: Hz
Options:
R&S
ZNA-K30
Manual operation:
See
[SENSe<Ch>:]NFIGure:HARMonic[:MAXimumorder]
<MaximumOrder>
Sets the sideband order of the sideband correction for a noise figure measurement in
channel <Ch>.
Suffix:
<Ch>
.
Channel number
Parameters:
<MaximumOrder>
Sets the number of higher order sidebands which are measured
and corrected during calibration and measurement.
*RST:
3
Options:
R&S
ZNA-K30
Manual operation:
See
[SENSe<Ch>:]NFIGure:IFConst
<IFFreq>
If manual selection is activated (
[SENSe<Ch>:]NFIGure:IFConst:STATe ON
command allows you to select an IF for niose figure measurements in channel <Ch>.
Suffix:
<Ch>
.
Channel number
Parameters:
<IFFreq>
Default unit: Hz
Options:
R&S
ZNA-K30
Manual operation:
See
[SENSe<Ch>:]NFIGure:IFConst:STATe
<Boolean>
Allows you to switch between automatic and manual IF selection for niose figure mea-
surements in channel <Ch>
Suffix:
<Ch>
.
Channel number
Parameters:
<Boolean>
OFF (0)
Automatic selection