Command reference
R&S
®
ZNA
1351
User Manual 1178.6462.02 ─ 20
.................................................................................... 1353
[SENSe<Ch>:]NFIGure:IFConst:STATe
.......................................................................... 1353
...................................................................................... 1354
.................................................................................... 1354
[SENSe<Ch>:]NFIGure:TEATtenuator
............................................................................1354
[SENSe<Ch>:]NFIGure:CALibration:NTIMe
<DetectorTime>
Defines the detector time of the RMS detector which is used for the noise measure-
ment during a NF calibration in channel <Ch>.
Suffix:
<Ch>
.
Channel number
Parameters:
<DetectorTime>
Detector time for the RMS detector
Default unit: s
Example:
SENSe1:NFIGure:CALibration:NTIMe 50ms
Options:
R&S
ZNA-K30
Manual operation:
See
[SENSe<Ch>:]NFIGure:CALibration:GTIMe
<DetectorTime>
Defines the detector time of the AVG detector which is used for the gain measurement
during a NF calibration in channel <Ch>.
Suffix:
<Ch>
.
Channel number
Parameters:
<DetectorTime>
Detector time for the AVG detector
Default unit: s
Example:
SENSe1:NFIGure:CALibration:GTIMe 10ms
Options:
R&S
ZNA-K30
Manual operation:
See
[SENSe<Ch>:]NFIGure:DEFine
<ReceivePort>, <DrivePort>
Initiates a noise figure measurement channel <Ch> and defines the receiving and driv-
ing port to which the DUT is connected.
Suffix:
<Ch>
.
Channel number
Setting parameters:
<ReceivePort>
Port number of the receiving port
<DrivePort>
Port number of the driving port
SCPI command reference