GUI reference
R&S
®
ZNB/ZNBT
788
User Manual 1173.9163.02 ─ 62
When using the NI-VISA library, ensure that the network analyzer itself is not listed as
a network device in the Measurement & Automation Explorer. Otherwise, "Scan Instru-
ments" will send an identification query (*IDN?), causing the analyzer to close the
"External Generators" dialog (without executing "Scan Instruments") and to activate the
remote screen.
Remote command:
n/a
LAN Detection
Activate "LAN Detection" to include the LAN interface in the autodetection sequence
(see
).
Note:
"LAN detection" will only work for external devices sharing the same IP subnet
with the R&S
ZNB/ZNBT.
As a prerequisite, the R&S
ZNB/ZNBT must accept incoming connections on UDP port
2473. If necessary, add a corresponding inbound rule to your firewall settings.
Remote command:
n/a
Add Device
Adds a new instrument to the list of "Configured Devices". In the "Add External Gener-
ator" dialog opened, you can specify the properties of the new instrument and the con-
nection.
●
"Interface" selects an interface/protocol type for the connection. In addition to the
GPIB, VXI-11, SOCKET, and USB-VISA interface types (for devices connected to
the GPIB Bus, LAN or USB connectors of the analyzer; see
), the ana-
lyzer supports any "Other" interface supported by the installed VISA library.
●
"Address" contains the address for the current interface type. GPIB addresses
must be unique for all devices connected to the GPIB bus (range: 0 to 30), GPIB
and IP addresses must agree with the entries in the VISA library. The remaining
interface types require composite address formats; see
If an instrument is connected to the R&S
ZNB/ZNBT, the entries in the [Driver for
New Instrument] panel can be auto-detected for the specified interface type and
address.
●
"Identify" sends an identification query ("IDN?") to the specified device address in
order to identify the type and serial number of the connected generator and select
Setup softtool