GUI reference
R&S
®
ZNB/ZNBT
789
User Manual 1173.9163.02 ─ 62
an appropriate driver file. Generator driver files (
*.gen
) are stored in the
Resources\ExtDev
subdirectory of the analyzer's program directory.
Table 5-9: Interface types for external generators and address formats
Physical
interface
(connector)
Interface
(protocol)
Address
Remarks
LAN
VXI-11
<IpAddress>
e.g.
127.0.0.0
Full VISA resource string:
TCPIP[board]::<Address>[::INSTR]
SOCKET
<IpAddress>::<PortNo>
e.g.
127.0.0.0::50000
LAN connection with pure TCP/IP protocol;
refer to your VISA user documentation.
GPIB
GPIB0 ...
GPIB9
<Address>
e.g.
20
Full VISA resource string:
GPIB[board]::<Address>[::INSTR]
USB
USB-VISA
<ManID>::<ProdID>::<SerialNo>
e.g.
0x0AAD::0x0047::100098
2733 (0x0AAD) is the manufacturer ID of
Rohde & Schwarz.
71 (0x0047) is an example for a R&S prod-
uct ID (R&S SMF100A).
The serial number is device-specific.
LAN or USB Other
Interface-specific, e.g. for SOCKET:
TCPIP0::<IpAddress>::<PortNo>::SOCKET
For USB-VISA:
USB0::<ManID>::<ProdID>::<SerialNo>::INSTR
Use complete VISA resource string.
Remote command:
SYSTem:COMMunicate:RDEVice:GENerator<Gen>:DEFine
5.18.4.5
Inline Calibration Systems dialog
USB-connected Inline Calibration Controllers (ICCs) are autodetected by the analyzer
firmware.
Setup softtool