Command reference
R&S
®
ZNB/ZNBT
1193
User Manual 1173.9163.02 ─ 62
7.3.14.4
[SENSe:]CORRection:CKIT...
The
[SENSe:]CORRection:CKIT...
commands deal with calibration kits and cal kit
data. The calibration kits are distinguished by their names (<CalkitName>), the optional
labels (<label>) can be used to carry information about the calibration standard.
In order to handle several identical calibration kits with different serial numbers use the
commands of
Chapter 7.3.14.5, "[SENSe:]CORRection:CKIT... with Labels"
[SENSe:]CORRection:CKIT:<StandardType>
.................................................................. 1193
[SENSe:]CORRection:CKIT:CATalog?
............................................................................ 1197
[SENSe:]CORRection:CKIT:DELete
............................................................................... 1197
.............................................................................. 1197
..................................................................................1198
[SENSe:]CORRection:CKIT:SELect
................................................................................1198
[SENSe:]CORRection:CKIT:<ConnType>:SELect
............................................................ 1199
[SENSe:]CORRection:CKIT:STANdard:CATalog?
............................................................ 1200
[SENSe:]CORRection:CKIT:STANdard:DATA?
................................................................ 1200
[SENSe:]CORRection:CKIT:<StandardType>
<ConnType>, <CalKitName>,
<StandardLabel>, <MinFreq>, <MaxFreq>[, <DelayParam>, <Loss>, <Z0>[,
<C0>, <C1>, <C2>, <C3>, <L0>, <L1>, <L2>, <L3>[, OPEN | SHORt | MATCh |
<Resistance>[, <Port1>[, <Port2>]]]]]
[SENSe:]CORRection:CKIT:<StandardType>?
<ConnType>, <CalKitName>[,
<Port1>[, <Port2>]]
Defines the parameters of a (possibly non-ideal) 1 port or 2-port calibration standard
<StandardType>
within a particular cal kit. Depending on the standard type, only a
subset of the parameters is used; see
Note
As the query does not reflect the delay mode specified using
, it is deprecated and only available for backward compatibility reasons.
Use
[SENSe:]CORRection:CKIT:STANdard:DATA??
instead.
Suffix:
<StandardType>
.
<string>
Standard type
For one-port standards, the first character denotes the gender,
for transmission standards the first two characters denote the
genders on both ends, e.g.
MOPen
for a male Open standard or
FFTHrough
for a Through standard with female connectors.
For a complete list of standards, refer to
.
Parameters:
<StandardLabel>
Additional string label for the standard, typically the standard's
serial number
SCPI command reference