THEORY OF OPERATION
5-2
A single chip microprocessor is used to control the operation of
the alarm buzzer, LED, pump and light sensor. It measures the
sensor readings and calculates the gas concentrations based
on calibration to known standard gases. The data are stored in
non-volatile memory so they may be sent to a PC for record
keeping. RS-232 transceivers provide a serial interface
between the monitor and the serial port of a PC. A two-line by
sixteen-character LCD is used to show the readings. The user
interacts with the monitor by using the [Y/+] key, [N/-] key, and
[MODE] key on the front membrane panel.
A rechargeable Lithium-Ion battery or a six C cell alkaline
battery pack powers the monitor.
Note:
The printed circuit board of the monitor is connected to
the battery pack even if the power is turned off. Therefore, it is
very important to disconnect the battery pack before servicing
or replacing sensors or any components inside the monitor.
Severe damage to the printed circuit board may occur if the
battery pack is not disconnected before servicing the unit.