PROGRAMMING OF IAQRAE
4-36
4.6.3 Enable/Disable Sensor
This function allows the user to selectively enable or disable
individual sensors in the IAQRAE monitor. When a sensor is
disabled, the sensor still measures the gas but no values are
displayed and all alarms for that sensor are rendered non-
functional.
1. When the “Enable/Disable Sensors?” screen appears,
press [Y/+]. The display will show all the sensors installed
in the monitor. An enabled sensor is marked with an
asterisk (*); a disabled sensor will have no asterisk. The
cursor will blink at the “OK?” message:
*CO
*VOC
*CO
2
*RH
OK?
*Temp
2. Press [MODE] to move from one sensor location to the
next one. Press [Y/+] to enable the sensor and [N/-] to
disable the sensor. Repeat until all the sensors that need
to be enabled have been selected.
3. Press [MODE] to move the cursor to “OK?” Press [Y/+] to
accept the change and move to the next submenu. Press
[N/-] to discard the change, and move to the next
submenu.