24
Programming Mode
Span Calibration VOC?
Refer to “Span Gas Calibration
for the PID Sensor” on page 13.
Zero Calibration?
Refer to “Zero Calibration for the
Oxygen Sensor” on page 14.
Calib Due Day?
Use this feature to set a reminder to
calibrate regularly. Enter the number of days between
calibrations per your company’s standards. After 30 days
of no calibration, the monitor sounds. Once the monitor
is calibrated, the counter resets.
Set Type?
Choose either automatic alarm reset or
latched (constant) alarm. To acknowledge and reset a
latched alarm, press
Y.
Set Run Silent?
Press
Y
to let the monitor run in silent
mode. In this mode, the audible alarm does not sound
during alarm conditions.
Press
N
to let the audible alarm sound during alarm
conditions.
Adjust Alarm Limits?
HIGH
- The highest reading for each gas concentration
since the monitor was turned on; updated every second.
Press
Y
to reset the high values.