RADWAG X2 Series Скачать руководство пользователя страница 7

- 7 - 

 

 
AS X2.PLUS balances 

 

 

 

1. power supply seat 
2. DB9/M connector (external buttons) 
3. COM 1 connector (e.g. printer) 
4. USB 2 type B connector (e.g. 

computer) 

5.

 USB 1 type A connector (e.g. 

computer keyboard)

 

6.

 RJ45 Ethernet connector

 

 

1.3. INTENDED USE 

X2 series balances are designed to provide accurate measurement of weighed loads, performed 
under laboratory conditions. 
 

1.4. PRECAUTIONS 

 

Prior to first use, it is highly recommended to carefully read this User Manual, and operate the 
balance as intended. 

 

Do not operate the touch panel using sharp-edged tools (knife, screwdriver, etc.). 

 

While loading the balance make sure that loads are placed in the very centre of the weighing 
pan. 

 

Load the weighing pan with loads, gross weight of which does not exceed instrument’s 
measuring range (maximum capacity). 

 

Do not leave heavy loads on the weighing pan for a longer period of time. 

 

In case of failure, immediately unplug the instrument from the mains. 

 

Balances to be decommissioned, should be decommissioned in accordance with valid legal 
regulations. 

 

Do not use the balance is areas endangered with explosion. The X2 series is not designed to 
operate in EX zones. 

 
NOTE: 
This equipment has been tested and found to comply with the limits for a Class B digital device, 
pursuant to part 15 of the FCC Rules.  These limits are designed to provide reasonable protection 
against harmful interference in a residential installation.  This equipment generates uses and can 
radiate radio frequency energy and, if not installed and used in accordance with the instructions, may 
cause harmful interference to radio communications.  However, there is no guarantee that 
interference will not occur in a particular installation.  If this equipment does cause harmful 
interference to radio or television reception, which can be determined by turning the equipment off 
and on, the user is encouraged to try to correct the interference by one or more of the following 
measures: 
- Reorient or relocate the receiving antenna. 
- Increase the separation between the equipment and receiver. 
-Connect the equipment into an outlet on a circuit different from that to which the receiver is 
connected. 
-Consult the dealer or an experienced radio/TV technician for help. 
 
 
NOTE:  
This device complies with part 15 of the FCC  rules.  Operation  is subject to the following two 
conditions:  

Содержание X2 Series

Страница 1: ...X2 Synergy SERIES AS X2 PLUS Analytical Balances PS X2 Precision Balances WLC X2 Precision Balances BRIEF USER MANUAL IMMU 16 S 03 03 21 EN www radwag com...

Страница 2: ...to be found on a CD delivered along with the device The full user manual version describes data crucial for settings and device operation If you are reading this it means that you are bound to achieve...

Страница 3: ...rature Stabilisation Time 16 3 7 Connecting Additional Hardware 16 4 Start Up 16 5 Keyboard Buttons Functions 17 6 Weighing Mode Home Screen 18 7 Operating Balance Menu 19 7 1 Entering Balance Menu 19...

Страница 4: ...4...

Страница 5: ...5 1 GENERAL INFORMATION 1 1 DIMENSIONS AS X2 PLUS series PS X2 series...

Страница 6: ...6 WLC X2 series 1 2 CONNECTORS X2 balances 1 Ethernet RJ45 connector 2 USB 1 Type A connector 3 USB 2 Type B connector 4 COM 1 connector 5 COM 2 connector 6 Power supply socket...

Страница 7: ...eas endangered with explosion The X2 series is not designed to operate in EX zones NOTE This equipment has been tested and found to comply with the limits for a Class B digital device pursuant to part...

Страница 8: ...nterference received including interference that may cause undesired operation NOTE The grantee is not responsible for any changes or modifications not expressly approved by the party responsible for...

Страница 9: ...ting or by electrical wiring failures defects caused by overloading the mechanical measuring system maintenance activities cleaning E Loss of warranty takes place if a repair is carried out outside RA...

Страница 10: ...free of vibrations and shakes free of air movement and dust Ambient air temperature should not exceed the range of 10 C 40 C Ambient relative humidity should not exceed 80 During balance operation amb...

Страница 11: ...bottom insert 2 centring ring embossment side up 3 weighing pan 4 draft shield 5 CAUTION AS balance with d 0 01 01mg is equipped with 2 weighing pans standard and open work Remember to calibrate the b...

Страница 12: ...ng of the balance level Level monitoring is performed uninterruptedly in the course of balance operation Level status is signalled with a respective pictogram displayed in the lower right corner of th...

Страница 13: ...is manual 4 Using a dry flannel cloth clean disassembled components mild cleanser may be applied if it does not contain any abrasive substances CAUTION Cleaning draft shield while still installed may...

Страница 14: ...n of other type could permeate into the panes thus causing weighing errors One shall not use driers when drying measuring glass tools It is a frequent treatment to leave glass components on a rack to...

Страница 15: ...the bottom insert Clean the weighing chamber and the panes All the operations have to be done carefully Pay special attention to the spot where the weighing pan is installed dirt and other small eleme...

Страница 16: ...iod thermal stabilisation period shall take at least 4 hours for PS and WLC balances and 8 hours for AS balances During the thermal stabilization the indications on a display panel can change It is re...

Страница 17: ...balance Press to Tare the balance Press to send measurement to a printer or a computer Function key Esc press to abandon parameter changes or exit to previous menu level Function key Home press to exi...

Страница 18: ...bol informing that communication with a USB flash drive is on Symbol informing that PC keyboard is connected Symbol informing that printer is connected via USB Symbol informing that communication with...

Страница 19: ...ection or clicking any button with particular parameter name results with change of colour This is for signalling purposes If a given area has any function or action assigned then it is performed auto...

Страница 20: ...measurement result can be saved or printed regardless of stability marker absence If unstable measurement result is printed then it is accompanied by question mark in front of printed mass value 8 1 G...

Страница 21: ...rises three permissions levels administrator advanced operator user Permissions level dependent access to edition of user parameters databases and software functions Permissions levels Enabled operati...

Страница 22: ...ue shall be displayed together with precise zero and stability markers Zeroing process is an equivalent for determining new zero point recognized by the balance as precise zero Zeroing is possible onl...

Страница 23: ...means of an internal adjustment weight button when pressed automatically triggers adjustment process Upon adjustment process completion respective message informing about process end and about its sta...

Страница 24: ...another parts assuming that mass of a single part is determined with sufficient accuracy and that the following parts are equal in mass Checkweighing Means of operation control of sample mass with app...

Страница 25: ...gs feature specific functions The functions enable adapting mode operation to your individual needs The special settings are activated on selecting a respective profile A detailed description of speci...

Страница 26: ...e 11 COMMUNICATION COMMUNICATION menu is comprised within Parameters menu It is accessed by pressing key The balance can communicate with a peripheral device wherein the communication is established v...

Страница 27: ...rement result press key Automatic printout is generated automatically in accordance with the settings for automatic printout The content of a printout depends on setting in menu Standard printout Weig...

Страница 28: ...28 14 ERROR MESSAGES...

Страница 29: ...29...

Страница 30: ...30...

Отзывы: