- 16 -
3.5. POWERING THE DEVICE
Balance can be connected to the mains only with a power adapter that comes standard with the
particular model. Nominal power supply of the power adapter (specified on the power adapter data
plate) should be compatible to the power from the mains.
Plug the balance to the mains – connect the power adapter to the socket, next connect its connector
to port located at the back of the balance housing.
Test of the display unit takes place right after connecting the balance to the power, all the elements
and pictograms are backlit for a short time. Next, the name and the program number appears, the
indication gets to ZERO (displayed reading unit depends on the balance). During the balance start,
the test of an internal mass adjustment mechanism occurs (single location and elevation of the
internal mass adjustment).
If the indication is different than zero, please press
button.
CAUTION! If the balance is “verified”, automatic adjustment occurs right after switching the balance on.
3.6. TEMPERATURE STABILISATION TIME.
Before start of measuring processes, it is necessary to wait until the balance reaches thermal
stabilisation.
For balances that were stored in much lower temperatures before plugging to mains (e.g. during
winter period), thermal stabilisation period shall take at least 4 hours for PS and WLC balances, and
8 hours for AS balances. During the thermal stabilization, the indications on a display panel can
change. It is recommended that ambient temperature changes at place of use were insignificant
(slow to change).
Thermal stabilization also applies to the level monitoring system of AS X2.PLUS balances.
3.7. CONNECTING ADDITIONAL HARDWARE
Use only accessories and peripheral equipment recommended by the manufacturer. The balance
must be disconnected from the mains before connecting or disconnecting any peripherals (printer,
PC computer, computer keyboard). On connecting the peripherals, plug the balance to the mains.
4. START-UP
•
Plug the power adapter to a socket, next connect the connector to port located at the back of
the housing.
•
Press
button
located
in the top right hand corner of
the terminal.
•
Wait until start-up procedure is completed, the home screen of balance software is displayed
automatically.
•
The balance runs with no user logged in. In order to start operation it is necessary to carry out
the logging procedure (for detailed logging procedure read later sections of this user manual).
CAUTION!
Remember to start the balance with no load on the weighing pan.
In accordance with the EN 45515 standard verified balances cannot display mass value below -20e.
If the indication value is below -20e, <Lo mass> information is displayed.
Press
key to zero the balance.