![RADWAG X2 Series Скачать руководство пользователя страница 20](http://html2.mh-extra.com/html/radwag/x2-series/x2-series_user-manual_3361296020.webp)
- 20 -
8. WEIGHING OPERATION
Load a weighed object on a balance weighing pan. On stabilization of weighing result, indicated by
stability marker
visible on the left side of balance display, read the measurement result.
Record / printout of the measurement result is available on pressing <PRINT> key:
•
for verified balances – only stable measurement result can be saved or printed (stability marker
visible on balance’s display),
•
for non-verified balances – stable or unstable measurement result can be saved or printed
(regardless of stability marker
absence). If unstable measurement result is printed then it is
accompanied by question mark <
?
> in front of printed mass value.
8.1. GOOD WEIGHING PRACTICE
In order to ensure long lasting use of a balance plus correct and reliable measurement of weighed
loads, follow below procedures:
•
Start the balance with no load on the weighing pan (permissible value of load on the weighing
pan on balance start is ±10% of its maximum capacity).
•
Load the weighing pan steadily avoiding shocks:
YES
NO
•
Place weighed loads centrally on the weighing pan:
YES
NO
•
Avoid side loading, in particular side shocks:
NO
NO