28. TROUBLESHOOTING
Problem Cause
Solution
Battery discharged.
Connect the scale to the mains,
charge the battery (batteries).
The scale does not switch
on.
No batteries (batteries not
installed or installed incorrectly).
Check if batteries are installed
correctly (polarization).
The scale switches off
automatically.
<7.3.t1> parameter set to value
enforcing scale shut-down after
particular time interval.
Go to 'Othr' menu, set <7.3.t1>
parameter to 'nonE' value.
During switching on,
message 'LH' is displayed.
Weighing pan loaded during
switching on process.
Unload the weighing pan. Zero
indication is displayed.