![RADWAG WLC/C2 Скачать руководство пользователя страница 20](http://html1.mh-extra.com/html/radwag/wlc-c2/wlc-c2_user-manual_3758479020.webp)
20
Options in case when the main unit is [g]:
•
g
(gram),
•
kg (kilogram),
•
ct (carat),
•
lb
(pound)*.
*)
–
unit disabled for verified weighing devices.
•
Select temporary unit and press
key, next go back to the home
screen.
12. ADJUSTMENT
option available for non-verified weighing devices exclusively
In order to ensure the highest weighing accuracy, it is recommended
to periodically introduce corrective factor of indications to scale memory, the
said factor must be referred to a reference weight. This is so called adjustment.
Adjustment has to be carried out:
•
prior
weighing,
•
if long breaks between successive measuring series occur,
•
if the ambient temperature has changed dynamically,
•
if the scale has been relocated.
Adjustment types:
•
external adjustment,
<1.1.CA-E>
, performed using an external weight
of declared mass, i.e. mass that cannot be modified,
•
user
adjustment,
<1.2.CA-u>
, performed using an external weight of mass
of any value comprised within the weighing range, however not lower than
30% of the maximum capacity value.
12.1. External Adjustment
External adjustment must be carried out using an external adjustment weight
of class F
1
.
Procedure:
•
Enter
<P1.CAL / 1.1.CA-E>
submenu, text
<UnLoAd>
(remove weight)
is displayed .
•
Remove the load from the weighing pan and press
key.