avoid side loads, especially side shocks:
10.2. Scales zeroing
In order to zero mass readout of a presently selected platform the user has
to press
button which is to be found in a weighment display of a
„MwManager”
program (the top right hand corner) or he has to initiate
zeroing function defined for a given input (See point
9.5.3 of this manual)
.
On a display mass readout with zero value will be shown next to symbols:
and
.
Zeroing means estimating a new zero point which will be recognized by scale
as an exact zero. Zeroing is possible only when display states are stable.
Caution:
Zeroing a display state is possible only within the range upto ±2% of the
maximum scales load. If zeroing value is greater than ±2% of the maximum
load then the message Err2 is displayed.
10.3. Taring
To estimate net mass for a presently selected platform the user needs to put
packaging on a device and wait for a readout to stabilize, next (s)he may
press button
. The other possibility is to initiate tare function defined
for a given input (See point
9.5.3 of this manual)
.
On a display mass readout with zero value is be shown next to symbols :
Net
i
. The scales has been tared.
Содержание PUE 5.15
Страница 2: ... 2 OCTOBER 2013 ...
Страница 7: ... 7 5 TERMINAL CONSTRUCTION 5 1 Overall dimensions Fig 1 PUE 5 15 terminal dimensions ...
Страница 8: ... 8 Fig 2 PUE 5 19 terminal dimensions ...
Страница 27: ... 27 Fig 13 Inputs Outputs configuration window ...
Страница 51: ... 51 Set chosen parameters and save the configuration to file see description in point 14 1 ...
Страница 55: ... 55 Fig 38 A D converter board Fig 39 Weighing module MW 04 ...