- 34 -
•
Dosing simulation
There are Stop dosing and Start dosing buttons in the bottom part of a
dosing simulation window. No matter what functions have been assigned
on Inputs, these buttons enable dosing process start and stop.
9.6.2. Checkweighing
Turning on
option and pressing
button opens
checkweighing settings window for a presently selected weighing platform.
Fig.22 Checkweighing parameters window
Description of the fields:
LO
threshold
Gross value of mass, checkweighing function is active if this
value is exceeded
Min
threshold
Mass value for estimation of tolerance threshold.
- MIN threshold value is signaled below the
Min threshold
value
- OK threshold is signaled between
Min threshold - Max
threshold
values.
- MAX threshold value is signaled over the
Max threshold
value
Max
threshold
Functions signalling within thresholds:
Содержание PUE 5.15
Страница 2: ... 2 OCTOBER 2013 ...
Страница 7: ... 7 5 TERMINAL CONSTRUCTION 5 1 Overall dimensions Fig 1 PUE 5 15 terminal dimensions ...
Страница 8: ... 8 Fig 2 PUE 5 19 terminal dimensions ...
Страница 27: ... 27 Fig 13 Inputs Outputs configuration window ...
Страница 51: ... 51 Set chosen parameters and save the configuration to file see description in point 14 1 ...
Страница 55: ... 55 Fig 38 A D converter board Fig 39 Weighing module MW 04 ...