- 20 -
Connection types
Selection of interface connection with a weighing module
RS 232
Connection via RS232 port
TCP/IP
Connection via Ethernet network
RS 485
Connection for RS 485 network
Offline
Offline mode is used for saving and edition of all indispensable
parameters. Saving and edition is carried out in a configuration file.
RS232:
Port
Choice of COM port number to which the module is physically connected
Speed
Transmission speed of RS232 communication port
By default
57600
bps
Parity
Parity state. By default „none” (not editable value)
Data bits
Number of data bits. By default 8 data bits (not editable value)
Stop bits
Number of stop bits. By default 1 stop bit (not editable value)
TCP/IP:
IP address
Device IP address, by default
192.168.0.2
Port
Port set in the weighing module, by default
4001
RS485:
Port
Choice of COM port number to which a module is physically connected
Speed
Transmission speed of RS485 communication port
By default
57600
bps
Parity
Parity state. By default „none” (not editable value)
Data bits
Number of data bits. By default 8 data bits (not editable value)
Stop bits
Number of stop bits. By default 1 stop bit (not editable value)
Address
Weighing module IP address
Caution:
1.
Weighing module MW-04 used in PUE 5 terminal communicates with an
internal PC module of the terminal via RS232 interface of COM6 port,
with the default transmission speed of 57600 bit per second
.
2.
Parameters <Connection settings> are not active in case of connection
with the weighing module.
Содержание PUE 5.15
Страница 2: ... 2 OCTOBER 2013 ...
Страница 7: ... 7 5 TERMINAL CONSTRUCTION 5 1 Overall dimensions Fig 1 PUE 5 15 terminal dimensions ...
Страница 8: ... 8 Fig 2 PUE 5 19 terminal dimensions ...
Страница 27: ... 27 Fig 13 Inputs Outputs configuration window ...
Страница 51: ... 51 Set chosen parameters and save the configuration to file see description in point 14 1 ...
Страница 55: ... 55 Fig 38 A D converter board Fig 39 Weighing module MW 04 ...