1. TECHNICAL DATA
AS series balances:
Type
AS
60/C/2
AS
60/220/C/2
AS
110/C/2
AS
160/C/2
AS
220/C/2
AS
310/C/2*
Maximum
capacity
60g 60/220g 110g 160g 220g 310g
Minimal load
1mg
1mg
10 mg
10 mg
10 mg
10 mg
Readability
0,01mg
0,01/0,1mg
0,1
mg
0,1mg 0,1mg 0,1mg
Tare range
-60g
-220g
-110 g
-160 g
-220 g
-310 g
Repeatability
0,02mg
0,02/0,1mg
0,1 mg
0,15 mg
0,15 mg
0,2 mg
Linearity
±0,05mg
±0,05/0,2mg
± 0,3 mg
Pan size
Φ
70 mm
Φ
85 mm
Sensitivity drift
2 ppm/°C in tempe15 ° - +35 °C
Working
temperature
+10 ° - +45 °C
Power supply
230V 50Hz AC / 11V AC
Calibration
Internal
(automatic)
Display LCD
with
backlight
* balance not verified
Type
AS 140/C/1/CT
AS 140/C/2/CT
AS 240/C/1/CT
AS 240/C/2/CT
AS 320/C/1/CT
AS 320/C/2/CT
Maximum
capacity
140g/700ct 240g/1200ct 320g/1600ct
Minimal load
0,004g/0,02ct 0,004g/0,02ct 0,004g/0,02ct
Readability
0,0002/0,001ct 0,0002g/0,001ct 0,0002g/0,001ct
Tare range
-140g/-700ct -240g/-1200ct -320g/-1600ct
Repeatability
0,0002g/0,001ct 0,0004g/0,002ct
Linearity
±0,0002g/0,001ct
±0,0004g/0,002ct
Pan size
Φ
85 mm
Sensitivity drift
2 ppm/°C in tempe15 ° - +35 °C
Working
temperature
+15 ° - +30 °C
Power supply
230V 50Hz AC / 11V AC
Calibration
internal (for C/2) or external (for C/1)
Display LCD
with
backlight