25
Troubleshooting and guide
Troubleshooting – when is it necessary?
Troubleshooting is necessary when the analyzer goes into
Operator Action Needed,
Troubleshooting needed
or
Intervention Required
mode. It may also be necessary to
troubleshoot messages in the
Analyzer status
screen.
About advanced troubleshooting
In the troubleshooting modes,
Troubleshooting needed
and
Operator Action Needed
, text and
video instructions will guide you through each troubleshooting procedure and show you what to do
to get out of the troubleshooting mode.
After each troubleshooting procedure, the analyzer tests to see if the issue has been resolved. If not,
a new troubleshooting procedure is shown on the screen. If the guided troubleshooting procedures
do not resolve the issue, the analyzer will go into
Intervention Required
mode.
To get out of Operator Action Needed mode
1. Follow the text and video instructions on the screen.
To get out of Troubleshooting needed mode
1. Follow the text and video instructions on the screen.
To get out of Intervention Required mode
1. Do the first action shown in the
Suggested actions
frame.
2. Tap the
Test again
button.
3. If the analyzer does not go out of
Intervention Required
mode,
do the next action.
4. Tap the
Test again
button.
5. If the analyzer does not go out of
Intervention Required
mode, do
steps 3 and 4 again.
6. If none of the actions cause the analyzer to go out of
Intervention
Required
mode, contact your local Radiometer service representative.
Please refer to the troubleshooting chapter "Possible causes", in the latest version of the
"ABL90 FLEX Instructions for use".
Note:
Those staff permitted to carry out the advanced troubleshooting must have
appropriate 'boxes' ticked when setting up their access rights under analyzer security.
25 of 36
EXT-VEND-MAN-0023-1 Owner:Julie Monnox Approved:27 Mar 2019