Changing consumables continued
Changing consumables – sensor cassette
Note:
The analyzer has to condition a new sensor cassette after it is
installed. This process can take up to 4 hours.
• Show on the analyzer screen how to enter Analyzer Status
and find > Consumables > Replace > Replace sensor cassette
• Show how to start the video guidance and wait until the sensor
cassette compartment opens
• Let the trainee try to remove the sensor cassette and dispose of it
as biohazardous waste according to local guidelines
• Tell that actions have to be confirmed at the analyzer to advance
the video guidance, so press > Action Completed
Let the trainee do all of the following steps:
• Pull the foil off the new sensor cassette pack
• Unscrew the lid and lift out the sensor cassette
• Observe that the trainee tap > Action Completed before putting
the new sensor cassette in place
• Follow the on-screen guidance for completing the procedure
Note:
If you tap the “Exit conditioning” button, the analyzer will not
perform the more frequent calibrations that normally follow in the 24
hours after a sensor cassette replacement. This may have a negative
effect on the accuracy and precision of patient results during this
time.