23
Replacements, continued
Consumables
Default tests or
Recommended replacement
activities per day
interval after installation
Solution pack
10
Maximum 30 days or when the
number
of
activities
is
zero
Sensor cassette
10
Maximum 30 days or when the
number
of
tests
is
zero
Inlet gasket holder
10
12 months
Inlet connector gasket
10
12 months
Can a sensor cassette be used again?
A sensor cassette removed from one analyzer can be used on the same or on another
ABL90 FLEX analyzer if the following six conditions are met:
• The sensor cassette is kept right side up after its removal. This prevents damage
to the sensors.
• The sensor cassette is installed within 2 hours of its removal
• The sensor cassette is installed before its Scheduled-to-replace date
• The sensor cassette is installed before its Expiration date
• The sensor cassette has some remaining tests
• The sensor cassette was not removed from an analyzer during a long-term shutdown
procedure
This data can be seen in the sensor cassette status screen.
Can a Solution Pack be used again?
A Solution Pack removed from one analyzer can be used on another if these 3 conditions
are met:
• the Solution Pack is installed before its
Scheduled to replace
: date
• the Solution Pack is installed before its
Expiration date
:
• the Solution Pack has some remaining activities
This data can be seen in the
Solution Pack Status
screen.
For further instructions, please refer to the latest version of “ABL90 FLEX Instructions for use”
(under “Replacement and maintenance”).
23 of 36
EXT-VEND-MAN-0023-1 Owner:Julie Monnox Approved:27 Mar 2019