Radiation Alert Inspector EXP+ Скачать руководство пользователя страница 8

8

3.  To select an option, push the “Set” button and the “SET” icon is displayed.  
4.  Use the “+” or “-” buttons to toggle between choices and press the “Set” button to enter the new value.  The

instrument will continue to operate in the Utility Menu mode, and the display will show "0" and "MENU".  To adjust

another Utility Menu option, repeat the above steps.

5.  To exit the Utility Menu at any time, press the SET button again.  The 

In continues with the

normal start-up routine.  

See options in the table below:

Options

Function

Comments

Interfacing to an External Device

The lower output jack (Figure 1(10)) on the left side of the Inspector

+

is a dual miniature jack that provides a data output

that can be used to drive a CMOS or TTL device. You can use it to record the counts on a computer, data logger, or
accumulating counter.  The output at the tip of the plug provides a positive (

10

 volt) pulse each time the Geiger tube detects

a count. Visit seintl.com for more information on Observer Software.

Options

WipeTest

Plate 

(patent # 5,936,246)

The stainless steel WipeTest Plate has a circular depression for placement of a wipe parallel to the detector window at a
fixed distance of 1 cm. The WipeTest Plate is designed to slide easily onto the back of the Inspector

+

.

4  Common Procedures

The following sections give instructions for several commonly-used procedures.  With any procedure, the user must
determine the suitability of the instrument or procedure for that application.

Establishing the Background Count

Normal background radiation levels vary at different locations, time, even in different areas of the same room.  To
accurately interpret the readings you get on the Inspector

+

, it is good to establish the normal background radiation count

rate for each area you plan to monitor.  You can do this by taking a timed count.  For more information on using the timer,
see “Taking a Timed Count” in Chapter 3.
A ten-minute average is moderately accurate.  You can repeat it several times and see how close the averages are.  To
establish a more accurate average, take a one-hour timed count.  If you need to determine whether there is prior
contamination, take averages in several locations and compare the averages.

Environmental Area Monitoring

You can keep the Inspector

+

in CPM or mR/hr mode whenever you want to monitor the ambient radiation, and look at it

from time to time to check for elevated readings.
If you suspect an increase in ambient radiation, use the timer and take a five or ten minute count, and compare the
average to your average background count.  If you suspect an increase that is too small to detect with a short timed
reading, you can take a longer count (for example 6, 12, or 24 hours).

Checking for Surface Contamination

CAUTION: 

Never touch the Inspector

+

to a surface that may be contaminated. You may contaminate the instrument. The

rubber strips on the back can be replaced if they become contaminated.  Replacement strips are supplied in this manual.

To check a surface, hold the detector window close to the surface and read the count rate (wait 30 seconds or until the
reading has stabilized).  If you want to find out if a surface is slightly radioactive, take a timed count or a longer
accumulated count.

1.

Auto Averaging or 3 sec. Averaging “on” selects Auto Averaging “oFF”

selects 3 second (fast response)
averaging

Refer to “Response Time
(Autoaveraging)” in Chapter 3

2.

Units Of  Measurements

Selects between mR/hr and CPM or 

µ

Sv and CPS

3.

Cal 100 Reset

Automatically resets Cal factor to 100

Press the set Button No toggling
required

4, 5, 6 

Reserved for future options

7.

Cal Factor Adjust 

Manually adjusts cal factor

Use “+” or “-” Buttons to increase or
decrease the value

8.

Factory Default Reset

Automatically resets to Auto Averaging,
mR/hr, CPM, and CAL 100

Press the set Button No toggling
required

9.

Revision #

Displays software version number

Содержание Inspector EXP+

Страница 1: ...User Manual...

Страница 2: ...lity Menu 7 Options 8 4 Common Procedures 8 Establishing the Background Count 8 Environmental Area Monitoring 8 Checking for Surface Contamination 8 5 Maintenance 9 Calibration 9 Calibrating Using a S...

Страница 3: ...spector in temperatures over 100 F 38 C or in direct sunlight for extended periods of time Do not get the Inspector wet Water can damage the circuitry and the mica surface of the Geiger tube Do not pu...

Страница 4: ...nel Each switch has three settings which are described below Mode Switch Figure 1 2 mR hr Sv hr The numeric display shows the current radiation level in milliroentgens per hour or when SI units are us...

Страница 5: ...et through the plastic case to penetrate the mica surface of the tube The small radiation symbol Figure 1 7 on the front label indicates the center of the Geiger tube External For Inspector EXP only T...

Страница 6: ...ility Menu detailed in Chapter 3 Refer to the following table After 30 second start up the reading will be based if instrument is detecting on an average of the previous 100 CPS 30 seconds 6000 CPM or...

Страница 7: ...e Using Dose Rate Modes While Timer is On Dose rate modes can be used while the timer is on In any dose rate mode the hour glass indicator will continue to flash during a timed period At the end of th...

Страница 8: ...e information on using the timer see Taking a Timed Count in Chapter 3 A ten minute average is moderately accurate You can repeat it several times and see how close the averages are To establish a mor...

Страница 9: ...nt down timer sound 11 Turn off the Inspector Hold down the button on the end panel and turn on the Inspector The display shows CAL and the Inspector counts down for 15 seconds chirping each second Th...

Страница 10: ...L and the hourglass icon flashing At the end of the 30 seconds it beeps The display shows CAL and SET is flashing 7 Press the and buttons to adjust the reading to what it should be When the reading is...

Страница 11: ...ta radiation A beta particle consists of an electron emitted from an atom It has more mass and less energy than a gamma ray so it doesn t penetrate matter as deeply as gamma and X rays Alpha radiation...

Страница 12: ...r 3 Operation Autoranging Display update CAL Factor Range 001 to 199 Timer Can set 1 10 minute sampling periods in one minute increments 10 50 minute sampling periods in 10 minute increments and 1 24...

Страница 13: ...or carrying the instrument Appendix C Inspector EXP Carrying Case The reinforced protective probe cover prevents damage to the fragile window of the detector The front flap lifts out of the way to gai...

Страница 14: ...m Observer Software The Observer software runs on a Windows platform and can be used with the Inspector Digilert 50 and Geiger Radiation Monitors As an option any of our Radiation Alert instruments ca...

Страница 15: ...onform to the warranty at any time while this warranty is effective the Warrantor will repair the defect and return the instrument to you prepaid without charge for parts or labor NOTE While the produ...

Страница 16: ...e company serial no Inside battery compartment or rear label address City state zip code 4 calibrations per year circle 1 2 3 4 phone number date placed in service Mail to Attn Robert Russell S E Inte...

Отзывы: