Taking a Timed Count

When a timed count is taken over a longer period, the average count per minute is more accurate, and any small increase
is more significant.  For example, if one 10-minute average is one count higher than another 10-minute average, the
increase may be due to normal variation.  But over 12 hours, a one-count increase over the 12-hour background average
may be statistically significant.
The Inspector

+

can give you a total count for a timed period from 1 minute to 24 hours. For a timed count of less than one

minute, watch the seconds countdown on the display. You can manually shut off the timer at any point.    

Follow these steps to take a timed count:

1. With the Inspector

+

turned on, set the Mode switch to 

Total/Timer 

and the timer switch on the end panel to

Set

. The

timer Hourglass and the SET icon appear on the display and the value is set to 00:01(1 minute).

2. Use the “+” and “-” buttons to adjust the timing period.  The timed period can be set for 1 to 10 minutes in one-minute

increments, for 10 to 50 minutes in ten-minute increments, or for 1 to 24 hours in one-hour increments. 

3. Set the Timer switch to 

On.  

The Inspector

+

beeps three times and starts counting.  The hourglass icon flashes during

the timed period.
If you want to see how many minutes remain, set the Timer switch to Set.  The display counts down from the time

setting in hours and minutes to zero.  For example, if the display says 00:21, 21 minutes remain.  During the timed

period, you can switch back and forth between 

Total/Timer

and the dose rate modes without interrupting the timed

period.  The hourglass indicator will show in any mode setting and will blink while the timer is totaling.

4. At the end of the timed period, the Inspector

+

beeps three times, and repeats the beeping several times over fifteen

seconds.  The number displayed is the total count.

5. Set the Timer switch to 

Off 

to return to normal operation.

To find the average counts per minute for the timed period, divide the total by the number of minutes.

6. To reset the timer to take another timed count, move the timer switch to the Off position and then to the On position.

The instrument will retain the timer setting of the last entered time.

Using Dose Rate Modes While Timer is On

Dose rate modes can be used while the timer is on.  In any dose rate mode, the hour glass indicator will continue to flash
during a timed period. At the end of the timed period, the hour glass will remain continuously on and the timed reading is
held in the Total/ Timer mode.

Taking a Total Count

The timer can take timed counts of up to twenty-four hours.  In certain situations, you may want to take a total count
without the timer; for example,  taking a count for longer than twenty-four hours.  Follow these steps:

1. Place the Inspector

+

in the location where you plan to take the count.

2. Note the time.
3. Immediately when you note the time, set the mode switch to 

Total/Timer.

4. At the end of the time period, note the time and the number of counts on the numeric display.
5. Subtract the starting time from the ending time to determine the exact number of minutes in the timing period.
6. To get the average count, divide the total counts by the number of minutes in the timing period.

Using The Alert

The Alert can be set in mR/hr or cpm. When using Utility Menu option #2 to switch units of measurements, the alert is
reset to the default alert level of .1 mR/hr (1.0 

µ

S/hr). Once the alert threshold is reached the beeper will sound until the

alert is deactivated, or the radiation level drops below the set alert threshold.

1.  To set the Alert, press the "SET" button on the end panel.  The "ALERT" icon (radiation symbol) and the "SET" icon

are displayed.  

2.  Use the "+" and "-" buttons to adjust the display to the desired level.  
3.  Press the "Set" button once to retain the setting in memory. Then press it  again to turn the alert mode on.  The

"ALERT" icon is now displayed to indicate the instrument is operating in the Alert mode.  

4.

To use the previous alert setting, press the set button twice. Now the alert mode is on.

5.  To deactivate the Alert mode, press the "Set" button again.  The "ALERT" icon is no longer displayed.

Utility Menu

The Utility Menu allows the user to change default settings for several operating parameters.  Once a setting is changed, it
remains in effect unless they are changed through the Utility Menu.  

1.  To activate the Utility Menu, hold down the “+” button while turning the instrument on.  The display will show the

"MENU" icon. Release the “+” button and a “1” indicating option 1 will appear along with the “Menu” icon.

2.  Scroll through the menu by pushing the “+” or “-” buttons.  

Содержание Inspector EXP+

Страница 1: ...User Manual...

Страница 2: ...lity Menu 7 Options 8 4 Common Procedures 8 Establishing the Background Count 8 Environmental Area Monitoring 8 Checking for Surface Contamination 8 5 Maintenance 9 Calibration 9 Calibrating Using a S...

Страница 3: ...spector in temperatures over 100 F 38 C or in direct sunlight for extended periods of time Do not get the Inspector wet Water can damage the circuitry and the mica surface of the Geiger tube Do not pu...

Страница 4: ...nel Each switch has three settings which are described below Mode Switch Figure 1 2 mR hr Sv hr The numeric display shows the current radiation level in milliroentgens per hour or when SI units are us...

Страница 5: ...et through the plastic case to penetrate the mica surface of the tube The small radiation symbol Figure 1 7 on the front label indicates the center of the Geiger tube External For Inspector EXP only T...

Страница 6: ...ility Menu detailed in Chapter 3 Refer to the following table After 30 second start up the reading will be based if instrument is detecting on an average of the previous 100 CPS 30 seconds 6000 CPM or...

Страница 7: ...e Using Dose Rate Modes While Timer is On Dose rate modes can be used while the timer is on In any dose rate mode the hour glass indicator will continue to flash during a timed period At the end of th...

Страница 8: ...e information on using the timer see Taking a Timed Count in Chapter 3 A ten minute average is moderately accurate You can repeat it several times and see how close the averages are To establish a mor...

Страница 9: ...nt down timer sound 11 Turn off the Inspector Hold down the button on the end panel and turn on the Inspector The display shows CAL and the Inspector counts down for 15 seconds chirping each second Th...

Страница 10: ...L and the hourglass icon flashing At the end of the 30 seconds it beeps The display shows CAL and SET is flashing 7 Press the and buttons to adjust the reading to what it should be When the reading is...

Страница 11: ...ta radiation A beta particle consists of an electron emitted from an atom It has more mass and less energy than a gamma ray so it doesn t penetrate matter as deeply as gamma and X rays Alpha radiation...

Страница 12: ...r 3 Operation Autoranging Display update CAL Factor Range 001 to 199 Timer Can set 1 10 minute sampling periods in one minute increments 10 50 minute sampling periods in 10 minute increments and 1 24...

Страница 13: ...or carrying the instrument Appendix C Inspector EXP Carrying Case The reinforced protective probe cover prevents damage to the fragile window of the detector The front flap lifts out of the way to gai...

Страница 14: ...m Observer Software The Observer software runs on a Windows platform and can be used with the Inspector Digilert 50 and Geiger Radiation Monitors As an option any of our Radiation Alert instruments ca...

Страница 15: ...onform to the warranty at any time while this warranty is effective the Warrantor will repair the defect and return the instrument to you prepaid without charge for parts or labor NOTE While the produ...

Страница 16: ...e company serial no Inside battery compartment or rear label address City state zip code 4 calibrations per year circle 1 2 3 4 phone number date placed in service Mail to Attn Robert Russell S E Inte...

Отзывы: