6.3
Troubleshooting
Mode
Alarming Indication
Situation
To do
Standby
Status Indicator:
X mark
AED will be turned
off after 8 to 10
times of electric
shock or 10 minutes
after turned on
Replace the
battery to a new
one
Audio message:
The battery is low,
please change the
battery immediately
Replace the
battery to a new
one
Audio message:
System error
Contact the vender
for A/S
AED is not turning
on when pressing the
power button.
Replace the
battery to a new
one, if the battery
is a new one, or
has already been
replaced
within 2 years,
please contact the
vender for A/S
Operating
Audio message:
The battery is low,
please change the
battery immediately
Replace the
battery to a new
one
Audio message:
System error
Contact the vender
for A/S
Audio message:
System will be
terminated due to no
attachment of the
pads
Attach AED pads
to the patient
properly
Attach AED pads
firmly to the skin
of the patient
Audio message:
Attach the pads to
the patient
Do not attempt to disassemble or repair the AED arbitrarily
Do not attempt to disassemble the AED arbitrarily due to danger of electric
shock hazard when high voltage is recharged. RADIAN is not responsible
for electric shock resulted from unauthorized disassembly or attempt to
repair. Please contact sales representative for defective AED