Chamber Energy Dependence
Specification Limits
Tube Voltage (kV)
Correc
tion Fac
tor
(with build-up)
1.10
0.90
1.00
0.95
1.05
10
20
50
100
200
500
1000
10X6-60E
Exposure Rate
200 nR/s - 2
R/s
2 pGy/s - 19
mGy/s
Dose
1
μ
R - 5.9 kR
10 nGy - 52 Gy
Cine (per frame) 0.01
μ
R - 100
mR
0.1 nGy - 1.0
mGy
Auto Dose
Threshold
216
μ
R/s
2
μ
Gy/s
Calibration
4% @ 50 kV, 0.88 mm Al HVL
Rate
Dependence
5%, 2 mR/min - 199 R/min
Energy
Dependence
5%, 0.2 mm Al HVL - 1.33 MeV
(with buildup material)
Construction
Parallel plate. 13 mm thick x 92
mm dia
60 cm
3
, 0.5 m cable
Application
Extended range. General purpose
for x-ray service
Use 8231 holder
32