Setup–Menu Functions
Pressing
select
at the main menu display
Change or SELECT
setup
enters
Setup
. The following information and options are provided in this mode::
!
Serial No. and firmware Version
This manual applies to version 6.
!
Pressure and temperature
Values used for ion-chamber compensation. Prefixed with Std if measured not selected.
!
SELECT restores setup defaults.
Restores all items in
Setup
to default values. Default values are shown as
Default
in the following
descriptions.
See detailed instructions on page 8
!
Mode SELECT
Parameter
Choices
Accu-Pro
Accu-Dose Accu-kV
Pressure
Measured
/Std 101.3 kPa
X
X
Temperature
Measured
/Std 22 deg C
X
X
Len/area suffix
On(*cm or *cm2)/
Off
X
X
High sensitivity
1
On/
Off
X
X
Radiation units
Roentgens
/Grays/Coulombs
X
X
Auto shutdown
On
(30 min)/Off
X
X
X
Time units
Seconds/
Minutes
/Hours
X
X
X
Mammo anode/filter
Mo/Mo,
W/Ag, W/Rh
X
X
DAC output
kV
/mA/Rel. intensity/dose rate
X
X
1
The Accu-Pro measures the background zero when you select a dose mode. If high sensitivity is set to
off, the system will make a quick coarse zero measurement. Some fluctuations will be observed in dose
rate mode and dose accumulate may exhibit drifting. If high sensitivity is set to on, the system will make
a fine zero measurement which takes 3 to 5 seconds. The dose rate will fluctuate much less and the
dose accumulate drift will be greatly reduced. It is recommended that high sensitivity mode is enabled for
most dose measurements.
!
Function SELECT
Add/Remove functions from the main menu. See Page 17 for a list of functions
!
kV disp(lay) SELECT
See table on page 12
!
Intrinsic filt(ration) SELECT
Selects the filtration used to compute kV in the diagnostic range. Auto enables automatic calculation
based on beam hardness.
Auto
, Cu (32 steps from 0 to 0.695 mm), Al (32 steps from 0 to 23 mm).