kV Dental
Operates like
kV Pulse
. Disregards low-intensity preheat pulses if this feature is enabled in
Setup
.
Duration can display in pulses for 50- and 60-Hz half-wave machines.
kV Fluoro
Operates like
kV Pulse
. Measures kV for 1 second, computes and displays the result and then repeats.
Displayed values change slightly to adapt to a continuous waveform. Width measurements are omitted;
dose and mAs become dose rate and average mA respectively.
Power off
Turns the unit off. Saves the operating mode and displays
Power off
for several seconds and then removes power.
Occurs automatically after 30 minutes of inactivity or when
power
is pressed during operation.