Warm-Up Cycle 

 

When you first power on the unit, the x-ray tube may or may not need to be 
properly warmed up (a warm-up is required every 40 hours or if there has been an 
electrical disruption). If the x-ray tube needs to be warmed up the system status 
box will say 

WARMUP REQUIRED

It is important to perform the warmup if this 

message appears. 

If the tube doesn’t need to be warmed up, the status box will say 

X-RAY OFF.  

To initiate a warm-up; 

 

1.

 

Press WarmUp on the main menu 

 

 

2.

 

The WARM-UPCYCLE menu will appear (see 
figure at the right). 

 

3.

 

There are two pre-programmed warm-up 
cycles labeled: 

NORMAL

 and 

EXTENDED

Unless you know that the unit hasn’t been 
used for a couple of months, select 

NORMALNORMAL 

 

4.

 

Press 

START 

(at the bottom of the screen). 

 

5.

 

The message in the white system status box 
will indicate 

WARMUP IN PROGRESS

. The 

flashing indicator lights on the unit will flash 
on and off indicating that x-rays are being 
produced. Leave the room for 15 minutes and 
note on the sign up sheet that a warm up was 
needed. 

6.

 

After the warm up is completed (theNORMAL 
warm-up cycle takes approximately 10 
minutes), that system status block will 
indicate 

WARMUP COMPLETE

. Press 

MAIN 

to 

take you back to the main menu. You are now 
ready to perform an irradiation which is 
discussed in the next section. 

 

  

 
 

Содержание RS-2000

Страница 1: ...egulatory Dose Limit General Hazards Associated with the X Ray Irradiator Safety Instructions Regarding the RS 2000 Irradiator Operating Instructions for the RS 2000 Irradiator 2 More complete and det...

Страница 2: ...ct with the body So our next question is What is ionization and why is it a concern All matter is made up of atoms Atoms contain a small central nucleus containing protons which have electrically posi...

Страница 3: ...called an ion The process of ionization is illustrated below This freed electron can potentially cause damage to living cells or DNA The damage can be repaired from small amounts of radiation however...

Страница 4: ...generate x rays see picture below Basically what happens is that a metal filament is heated up to a very high temperature allowing electrons from the metal to essentially boil off Electrons the small...

Страница 5: ...ill I receive from handling irradiated specimens and from using this irradiator As for handling irradiated specimens they do NOT become radioactive after an irradiation and you will not receive any ra...

Страница 6: ...s No unauthorized personnel may defeat or override any safety features on the x ray generator collimator or shielding without permission of the manufacturer Rad Source The RS 2000 is to be serviced by...

Страница 7: ...Indication Lights show when X rays are being produced Shielded and Interlocked X Ray Tube Chamber to prevent X ray production if panel should be opened Door Handle Double Safety Interlocked to prevent...

Страница 8: ...ning module must be documented in order to get Xray access This training is different that the Radiation training orientation you need to complete both Radiation Safety for Radioactive Material Users...

Страница 9: ...t Here are the different steps that may need to be taken Let me know which one fits your needs and I will send you the correct forms 1 If you currently have a mouse protocol and have mice housed in CC...

Страница 10: ...gaged unit will not operate if not properly closed This can be done after you load your samples animals and set your time 2 Use the key to turn the power on off switch to on the key to power on the un...

Страница 11: ...d that over heating can cause the unit to shut down I also have an electric fan in the room that you can set up to bring cooler air from the hall into the room 5 Main Menu The main menu enables the op...

Страница 12: ...re two pre programmed warm up cycles labeled NORMAL and EXTENDED Unless you know that the unit hasn t been used for a couple of months select NORMALNORMAL 4 Press START at the bottom of the screen 5 T...

Страница 13: ...re typically irradiated on Level 1 i e within the black RAD reflective chamber on the floor of the chamber Place no more than 5 mice in a CLEAN cage Remove food and water from the cage if you have exc...

Страница 14: ...ormity is 98 The radiation dose rate is highly dependent upon the distance from the radiation source x ray tube Therefore for certain specimens e g test tubes flasks plates you should lay items horizo...

Страница 15: ...S 29M 42S 22M 43S 16M 40S 11M 43S 7M 19S 3 500 35 31M 15S 30M 42S 34M 39S 26M 32S 19M 27S 13M 41S 8M 33S 4 000 40 35M 42S 35M 06S 39M 36S 30M 18S 22M 14S 15M 38S 9M 45S 6 000 60 53M 35S 52M 32S 59M 24...

Страница 16: ...an irradiation time in for one programs A B C or D If you have already programmed in an irradiation time go to Step 3 After selecting PROGRAM SETUP the following screen AUTO PROGRAM SETUP will appear...

Страница 17: ...afety Cabinet If the BSC is used spray and wipe down all surfaces inside the BSC Repeat with TEC Surf solution c Please dispose of all debris and waste in the waste basket provided d Empty the bleach...

Страница 18: ...rrective actions for the RS 2000 Most of the time you will just need to hit clear and continue on If that doesn t work turn off the machine wait 20 seconds and turn it back on If neither of those opti...

Отзывы: