3151 And 3151A User Manual
Using The Instrument 3-23
Selecting the
Trigger Slope
The trigger slope command selects the sensitive edge of the trigger
signal that is applied to the TRIG IN connector. The Model
3151/3151A can be made sensitive to either positive or negative
transitions. Use the following command to select the edge sensitivity
for the trigger signal:
TRIGger:SLOPe {POSitive | NEGative}
Positive going transitions will trigger the Model 3151/3151A when the
POS option is selected. Negative transitions will trigger the Model
3151/3151A when the NEG option is selected. POS is the default
slope.
The query:
TRIGger:SLOPe?
queries the trigger slope and returns POS or NEG.
Using the Soft
Trigger
There are a number of commands that are available to trigger the
Model 3151/3151A. The soft trigger command is one of them. To
use the soft trigger command, place the instrument in the
TRIG:SOUR EXT mode. Soft trigger is ignored in the internal or
TTLTrg<n> modes. Use the following SCPI commands to trigger the
instrument:
TRIGger
The IEEE-STD-488.2 common command *TRG will have the same
effect. Use either software command to trigger the Model
3151/3151A in Trigger, Burst and Triggered Sequence Advance
modes.
Содержание 3151
Страница 4: ...This page was left intentionally blank...
Страница 24: ...3151 And 3151A User Manual Getting Started 1 12 This page was left intentionally blank...
Страница 29: ...3151 And 3151A User Manual Configuring The Instrument 2 5 Figure 2 1 Set The Logical Address...
Страница 74: ...3151 And 3151A User Manual Using The Instrument 3 44 This page was left intentionally blank...
Страница 98: ...3151 And 3151A User Manual SCPI Command Reference 4 24 This page was left intentionally blank...
Страница 126: ...3151 And 3151A User Manual Optional Harness Assemblies 6 2 This page was left intentionally blank...
Страница 127: ...3151 And 3151A User Manual Optional Harness Assemblies 6 3...
Страница 131: ...3151 And 3151A User Manual Optional Harness Assemblies 6 7...