3151 And 3151A User Manual
Maintenance and Performance Checks 5-5
Recommended Test
Equipment
Recommended test equipment for troubleshooting, calibration and
performance checking is listed below. Test instruments other than
those listed may be used only if their specifications equal or exceed
the required characteristics.
Test Instrument & Accessories
Minimum Specification
Universal Counter
100 MHz, 10 ppm accuracy
DMM
ACV, 0.2%; DCV, 0.1% accuracy
Distortion Analyser
10 Hz to 100 kHz
Spectrum Analyser
1 GHz bandwidth, 1 kHz resolution
Pulse/Function Generator
20 MHz, 10 ns pulsewidth
Oscilloscope
2 channels, 400 MHz analog bandwidth
Feedthrough termination
50
Ω
, 1%
BNC Cables
1 meter long, 50
Ω
impedence
20dB feedthrough attenuator
50
Ω
, 1%
Performance Check
Procedures
Use the following procedures to check the Model 3151/3151A
against the specifications. A complete set of specifications is listed
in Appendix A. The following paragraphs show how to set up the
instrument for the test, what the specifications for the tested function
are, and what acceptable limits for the test are. If the instrument
fails to perform within the specified limits, the instrument has to be
calibrated or tested to find the source of the problem.
Содержание 3151
Страница 4: ...This page was left intentionally blank...
Страница 24: ...3151 And 3151A User Manual Getting Started 1 12 This page was left intentionally blank...
Страница 29: ...3151 And 3151A User Manual Configuring The Instrument 2 5 Figure 2 1 Set The Logical Address...
Страница 74: ...3151 And 3151A User Manual Using The Instrument 3 44 This page was left intentionally blank...
Страница 98: ...3151 And 3151A User Manual SCPI Command Reference 4 24 This page was left intentionally blank...
Страница 126: ...3151 And 3151A User Manual Optional Harness Assemblies 6 2 This page was left intentionally blank...
Страница 127: ...3151 And 3151A User Manual Optional Harness Assemblies 6 3...
Страница 131: ...3151 And 3151A User Manual Optional Harness Assemblies 6 7...