3151 And 3151A User Manual
Maintenance and Performance Checks 5-9
Squarewave
Characteristics
Specified Transition Time: <6 ns (10% to 90% of amplitude)
Specified Aberration: <5% of amplitude
Equipment: Oscilloscope, 20dB attenuator
1. Configure the Model 3151/3151A as follows:
Function:
Square wave
Frequency:
1.000 MHz
Amplitude:
10 V
Remote Commands:
:RES
:OUTP ON
:FUNC:SHAP SQU
:FREQ 10e6
:VOLT 10
2. Connect the Model 3151/3151A output to the oscilloscope input.
Use the 20 dB attenuator and set the oscilloscope input
impedance to 50
Ω
.
3. Set the oscilloscope and verify that the rise and fall times are less
than 6 ns. Verify that overshoot and ringing are less than 5%
of amplitude.