User Manual 1260-60A/B
Module Specific Syntax 3-1
Chapter 3
MODULE SPECIFIC SYNTAX
1260-60A/B
Syntax
The Module Specific Syntax for the 1260-6A,B module is as
follows:
<mod addr>.<bank no><relay no>
where <mod addr> is the address of the 1260-60A,B.
NOTE
The <mod addr> used here is NOT the VXIbus
defined logical address of the 1260 Series Master. It
is peculiar to the 1260 Series and describes the
switching module in relation to the 1260 Master. This
address corresponds to the binary value of the
switch setting of SW1 on the switching module PCB.
<bank no> is a reference to the bank of the relay to be switched.
It is a single digit number.
The <bank no> refers to the following relay banks:
0
1x12BankA
1
1x12BankB
2
1x2 Relays S1, S2, S3
<relay no> refers to the relay to be operated. This is a two-digit
number. For Bank A and Bank B, this value must be between 00
and 11. For relay S1 it is 00, for S2-01, and for S3-02. Note the
leading 0 for relays 00 through 09 is required.
Refer to Figures 4-1, 4-2, and Table 4-1 for banks, relay
numbers, and connector pins for the 1260-60A,B module.
If more than one connection is to be made or broken on the
1260-60A,B with contiguous relays, the following format is
supported:
<mod addr>.<bank no><relay no>-<bank no><relay no>
Multiple groups of relays can be specified on a single command