User Manual 1260-60A/B
Module Specific Syntax 3-4
sending the command:
CNF ON
2. Each time a relay is opened or closed, ensure the operation
is complete using the following procedure:
a. Send the OPEN or CLOSE command, e.g.:
OPEN 3.200
b. Send the YERR command to read back the error
YERR
c. Read the error result. The error result should be:
error 000.00