User Manual 1260-30
Installation Instructions 2-6
1260-30D
The following connections are required to configure the 1260-30
as a 1260-30D. To configure the 1260-30B from any other 1260-
30 configuration, remove all connections not on this list and install
22 gauge solid jumpers, listed below, unless otherwise specified.
*COM 01 to E00-1
Teflon stranded, 24 gauge, white
COM 02 to E00-2
Teflon stranded, 24 gauge, black
*COM 11 to E10-1
Teflon stranded, 24 gauge, white
COM 12 to E10-2
Teflon stranded, 24 gauge, black
*COM 21 to E20-1
Teflon stranded, 24 gauge, white
COM 22 to E20-2
Teflon stranded, 24 gauge, black
*COM 31 to E30-1
Teflon stranded, 24 gauge, white
COM 32 to E30-2
Teflon stranded, 24 gauge, black
*COM 41 to E40-1
Teflon stranded, 24 gauge, white
COM 42 to E40-2
Teflon stranded, 24 gauge, black
*COM 51 to E50-1
Teflon stranded, 24 gauge, white
COM 52 to E50-2
Teflon stranded, 24 gauge, black
*COM 61 to E60-1
Teflon stranded, 24 gauge, white
COM 62 to E60-2
Teflon stranded, 24 gauge, black
*COM 71 to E70-1
Teflon stranded, 24 gauge, white
COM 72 to E70-2
Teflon stranded, 24 gauge, black
W1-1 to W1-2
* Twisted Pair
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com