Use “?” to list available commands as shown below.
>?
YOUNG 61302L SERIES BAROMETER V2.02
----------------------------------------
CMD100 n OUTPUT PERIOD (0-60 sec)
CMD105 n AVG SAMPLE COUNT (0-32)
CMD110 n OUTPUT FORMAT
(1=ASCII 2=POLLED 3=NMEA 4-RAW 5-DIAG)
CMD120 n MULTIPLIER (10000 = 1.0000)
CMD125 n OFFSET (hPA x 100)
CMD130 n SEA LEVEL CORRECTION (hPA x 100)
CMD140 n IOUT RANGE HI (500-1100)
CMD142 n IOUT RANGE LO (500-1100)
CMD150 c POLL ADDR CHAR
CMD160 n BAUD RATE (38400,9600,4800,2400,1200)
CMD300 n 1=STOP, 2=RUN
CMD310 n FORCE IOUT (0-20000 FULL SCALE)
CMD400 n 1=RPT TEXT, 2=RPT CODE ONLY
CMD410 GET CAL TABLE
CMD420 STORE SETUP PARAMETERS
>
CMD100 n
sets output period in seconds. When set to
zero, the output rate is 1.8 per second.
CMD105 n
sets the average sample count. Higher sample
count yields more stable readings but will lengthen
the startup delay.
CMD110 n
sets output format.
CMD120 n
sets the output multiplier. 10000 represents
1.0000
CMD125 n
sets the output offset in hPa x 100
CMD130 n
sets the sea level correction in hPa x 100
=[
P
unadjust
*(
Mult / 10000
)]
+
[(
Offset
/
100
)
+
(
Correction
Sea Lvl
/
100
)]
CMD140 n
sets the IOUT Pressure Range HI
CMD142 n
sets the IOUT Pressure Range LO
IOUT Pressure Range HI must be greater than IOUT
Pressure Range LO
CMD150 c
sets the poll address character. Any ASCII
character from ‘0’ (ASCII 48) to ‘z’ (ASCII 122) may
be used.
CMD160 n
sets the baud rate. Acceptable values are
38400, 9600, 4800, 2400, and 1200. Note that the
baud rate change occurs immediately after the
command is received. Therefore, in order to continue
communicating, the serial communication program
that issued the command must also change to
match.
CMD300 n
causes the sensor to pause or resume
measurement. 1 = Stop, 2 = Run.
CMD400 n
causes the sensor to issue a report
summarizing parameter settings.
1 = Text (parameter values and labels)
2 = Code Only (parameter values only).
CMD420
burns the current parameter settings to flash
memory. This step
must
take place to retain
changes to settings otherwise parameters will revert
to previous values at the next power up.
5.0 MAINTENANCE
The MODEL 61302L barometer requires no regular
maintenance. Periodic calibration certification, if required,
is available from the factory.
6.0 WARRANTY
This product is warranted to be free of defects in materials
and construction for a period of 12 months from date of
initial purchase. Liability is limited to repair or replacement
of defective item. A copy of the warranty policy may be
obtained from R. M. Young Company.
7.0 EMC COMPLIANCE
This device complies with Part 15 of the FCC Rules. Operation
is subject to the following two conditions: (1) this device may
not cause harmful interference, and (2) this device must accept
any interference received, including interference that may cause
undesired operation.
This equipment has been tested and found to comply with the
limits for a Class A digital device, pursuant to part 15 of the
FCC Rules. These limits are designed to provide reasonable
protection against harmful interference when the equipment
is operated in a commercial environment. This equipment
generates, uses, and can radiate radio frequency energy and, if
not installed and used in accordance with the instruction manual,
may cause harmful interference to radio communications.
Operation of this equipment in a residential area is likely to
cause harmful interference in which case the user will be
required to correct the interference at his own expense.
This ISM device complies with Canadian ICES-001.
Cet appareil ISM est conforme à Ia norme NMB-001 du Canada.
EN55011/CISPR 11, Group 1, Class B device.
Class B equipment is suitable for use in domestic establishments
and in establishments directly connected to a low voltage power
supply network which supplies buildings used for domestic
purposes.
8.0 CE COMPLIANCE
This product has been tested and complies with European
CE requirements for the EMC Directive. Please note that
shielded cable must be used.