13
DIFFERENCES USING A QUPSI
13
Differences using a quPSI
The same software realizations and user library can be used with a quPSI device. Just
be sure to take note of the following differences.
13.1
quTAU GUI
When using the quTAU GUI with a quPSI device, please note that external countrates are
detected in channel 1 (Sync), 3 and 4, while the internal APDs countrates are displayed
in channel 5 and 6, see also Table
. There is also a fixed sync divider value of 1024 for
channel 1. In the settings window, only the first 3 rows are used to change input settings
for the external sources.
Table 12:
Channel Mapping of the quPSI device
. Please note that this only applies for
read out timestamps and count rates, but not for the input settings.
Source
External, Sync
External
Internal APDs
Channel
1
3 & 4
5 & 6
In the HBT tab of the GUI, there is now the option to switch between internal and
external APDs for the analysis (instead of being able to freely choose channels).
13.2
DLL Usage
Please note that the function
TDC setHbtInput
used to set the channels for HBT-analysis
will return an error when used with a quPSI device. Instead, use
TDC switchHbtInternalApds
to switch between internal and external APD usage. Other than that, functions will be-
have similarly for the quTAU and quPSI devices.
quTAU/quPSI Manual V4.0 Page 44